beautypg.com

Metrohm 726 Titroprocessor User Manual

Page 151

background image

88. . Command Reference

Command Reference

726 Titroprocessor, Instructions for use

145

145

Signal drift

Limits for signal drift

with MET_PH/U/IPOL:

0,5…50…999 mV/Min | off

with MET_UPOL:

0,1…50…100 µA/Min | off

After adding a volume step, the next addition of reagent

is not made until the change in the measured value of

the sensor falls below the indicated value for the signal

drift. If this is delayed, dosing is continued irrespective

on expiry of the pre-set equilibration time (see be-

low).This signal drift control can be switched off ('off’).

When the drift control is switched off, the pre-set waiting

time (see below) is strictly adhered to. This can be use-

ful if the titration reaction proceeds slowly or the sensor

takes a long time to respond.

Equilibr.time

max. waiting time before next volume step

0,1…9999 s | auto | off

The full waiting time between one volume increment and

the next volume step takes effect when, with the drift

control switched on (see signal drift, above), the signal

drift does not fall below the pre-set value. Otherwise,

dosing is continued after attainment of the signal drift.

The full waiting time also takes effect if the drift control is

switched off.

The equilibration time function can be switched off. In

this mode, the only decisive factor is the signal drift for

the titration procedure.

In the 'auto' mode the Titroprocessor selects an appro-
priate equilibration time in line with the pre-set signal

drift value. It is calculated using the following formula:

Equilibration time in s

drift mV

(

)

/ (

/ min)

,

=

+

+

150

0 01

5

For the standard 50 mV setting this results in a 26 sec-

ond equilibration time.

8.3.3 EP Recognition

If there is a possibility of several equivalence points occurring in one titra-

tion, it is essential for a correct evaluation that the criteria for recognizing the

EPs be correctly set.

The parameters of EP recognition for the montonic titration modes coincide

with those of the DET modes, see page 138ff.

As the ERC is calculated differently for a monotonic titration, this results in a

different range of validity for the EP criterion, see next page.