22. . Installation

Installation

726 Titroprocessor, Instructions for use

25

25

2.6.5 Sensors / Electrodes

1

Ind A2 Ind A1

Ref

Pol

Pt100/1000

Sensors A

Titroprocessor models 2.726.0110 and 2.726.0120 are supplied with two

built-in measuring interfaces (Sensors A and Sensors B). The models

2.726.0010 and 2.726.0020 only have one measuring interface (Sensors A)

but can be extended to two measuring interfaces at a later date with

upgrade set 6.5855.000.

The different measuring inputs of a measuring interface:

Ind A1, A2

Input for indicator or measuring electrodes (single and combined elec-

trodes) for potentiometric determinations.

It is possible to carry out a manual measurement at one measuring input

while the second input is carrying out a titration.

Ref

Separate reference electrode input. If combined electrodes are used this

electrode input is not required.

Pol

Connection for polarized electrodes for amperometric and voltametric

determinations (measuring modes Upol or Ipol).

Pt100/1000

Inputs for Pt100 or Pt1000 temperature sensors (also used for electrodes

with built-in Pt100 or Pt1000 sensors).

Stirrer connection

Stirrer power supply (12 V, 250 mA) for operation without sample

changer. Can be used with 728 Magnetic stirrer, 722 Rod stirrer and 727

and 703 Titration stands.

The measuring inputs Ref, Ind A1 and Ind A2 (or B1, B2) can be used to-

gether as a differential amplification switch for differential potentiometry. This

is recommended for measurements or titrations in low-conductivity media

(e.g. organic solvents).