beautypg.com

Metrohm 726 Titroprocessor User Manual

Page 147

background image

88. . Command Reference

Command Reference

726 Titroprocessor, Instructions for use

141

141

Fix end points

End points with pre-set measured values

It may be desirable to calculate with a volume value at a

fixed measured value on the titration curve. This is

where fix end points can be useful.

F# M.value

Measured value of the fixed end point

with DET_PH*

pH -20…20 | off

with DET_U*, DET_IPOL*

-2000…2000 mV | off

with DET_UPOL*

-200 …200 µA | off

Up to nine fix end points can be defined. Upon titration,

their associated volume values are filed in the F1…9 va-

riables (measured values of the fix end points =>

FM1…9) and can be used for calculations like normal

EPs.

8.2.4 Curve display

The settings for the curve display are valid for both the live curve display

and the curve printout. The measurable variable display for the two axes and

their scales can be set individually. Combined curves can also be displayed

(not live).

Quantity Y_Axis

Measurable variables on the Y-axis

M.value | Volume | Temp. | dMW/dV

MW,Temp. | MW,dMW/dV | all

Other display options are possible in addition to the

standard Measured Value/Volume display. The choice

of measurable variables on the Y-axis is as follows:

M.value

pH, mV or µA (in line with mode)

Volume

i.e. volume against time

Temp.

with automatic measurement only

dMW/dV

1st derivative of the measured

value curve against volume

MW, Temp.

Meas. value and temp. combined

MW,dMW/dV

Meas. value and 1st derivative

against volume, combined

all

all displayable curves combined