Metrohm 726 Titroprocessor User Manual
Page 56
3.4 Buret
Buret units
726 Titroprocessor, Instructions for use
50
50
3.4.3 685 Dosimat, exchange units
For the 685 Dosimat exchange units the types of buret units are not defined
as tubing dimensions. In this case the rinsing volume and the number of
rinsing cycles are entered .
The same names can be used as for the dosing units of the 700 Dosino.
The 726 Titroprocessor can determine the type of dosing unit itself.
ÚÄÄÄConfiguration ДДДДДДДДДДДДДДДДДДДДДДДДД(Program version 5.726.0012)ДДДДДДДД¿
³ ³
³ DialÚÄÄÄ
Tubings
ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿ ³
³ ³ ³ ³
³ LCD ³ Dosimat 685: Exchange units, preparation ³ ³
³ LCD ³ ³ ³
³ ³ # Type Volume Cycles Dosing rate Filling rate ³ ³
³ Fehl³ ДД ДДДДДДДД ДДДДДДДДДД ДДДДДД ДДДДДДДДДДДДД ДДДДДДДДДДДДД ³ ³
³ Edit³ 1 Default 3.5 mL 2 max mL/min max mL/min ³ ³
³ Edit³ 2 Changer 10.0 mL 2 max mL/min max mL/min ³ ³
³ ³ 3 mL mL/min mL/min ³ ³
³ ³ 4 mL mL/min mL/min ³ ³
³ ³ 5 mL mL/min mL/min ³ ³
³ ³ 6 mL mL/min mL/min ³ ³
³ ³ 7 mL mL/min mL/min ³ ³
³ ³ 8 mL mL/min mL/min ³ ³
³ ³ 9 mL mL/min mL/min ³ ³
³ ³ 10 mL mL/min mL/min ³ ³
³ ³ ³ ³
³ ³ ³ ³
³ ³ ³ ³
³ АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ ³
³ ³
АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ
ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿
³ ³ ³ ³ ³ ³ ³ 700 ³ ³Delete ³Delete ³ Check ³
³ ³ ³ ³ ³ ³ ³Dosino ³ ³line ³ all ³ type ³
АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ
Type
Name of the buret unit. Two types are already predefined.
(Default=standard settings, see above,
Changer
=standard setting for sample changer
applications, see above.)
Volume
Reagent volume for one rinsing cycle.
Cycles
Number of rinsing cycles to be carried out.
Dosing
rate
The maximum value for the dosing speed depends on the
cylinder size of the buret used.
Buret cylinder
Max. speed
1 mL
3 mL/min
2 mL
6 mL/min
5 mL
15 mL/min
10 mL
30 mL/min
20 mL
60 mL/min
50 mL
150 mL/min
For smaller buret cylinders a value of up to 150 mL/min
can also be entered, but when the function is carried out