beautypg.com

Metrohm 726 Titroprocessor User Manual

Page 56

background image

3.4 Buret

Buret units

726 Titroprocessor, Instructions for use

50

50

3.4.3 685 Dosimat, exchange units

For the 685 Dosimat exchange units the types of buret units are not defined

as tubing dimensions. In this case the rinsing volume and the number of

rinsing cycles are entered .

The same names can be used as for the dosing units of the 700 Dosino.

The 726 Titroprocessor can determine the type of dosing unit itself.

ÚÄÄÄConfiguration ДДДДДДДДДДДДДДДДДДДДДДДДД(Program version 5.726.0012)ДДДДДДДД¿

³ ³

³ DialÚÄÄÄ

Tubings

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿ ³

³ ³ ³ ³

³ LCD ³ Dosimat 685: Exchange units, preparation ³ ³

³ LCD ³ ³ ³

³ ³ # Type Volume Cycles Dosing rate Filling rate ³ ³

³ Fehl³ ДД ДДДДДДДД ДДДДДДДДДД ДДДДДД ДДДДДДДДДДДДД ДДДДДДДДДДДДД ³ ³

³ Edit³ 1 Default 3.5 mL 2 max mL/min max mL/min ³ ³

³ Edit³ 2 Changer 10.0 mL 2 max mL/min max mL/min ³ ³

³ ³ 3 mL mL/min mL/min ³ ³

³ ³ 4 mL mL/min mL/min ³ ³

³ ³ 5 mL mL/min mL/min ³ ³

³ ³ 6 mL mL/min mL/min ³ ³

³ ³ 7 mL mL/min mL/min ³ ³

³ ³ 8 mL mL/min mL/min ³ ³

³ ³ 9 mL mL/min mL/min ³ ³

³ ³ 10 mL mL/min mL/min ³ ³

³ ³ ³ ³

³ ³ ³ ³

³ ³ ³ ³

³ АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ

ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿

³ ³ ³ ³ ³ ³ ³ 700 ³ ³Delete ³Delete ³ Check ³

³ ³ ³ ³ ³ ³ ³Dosino ³ ³line ³ all ³ type ³

АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ

Type

Name of the buret unit. Two types are already predefined.
(Default=standard settings, see above,
Changer

=standard setting for sample changer

applications, see above.)

Volume

Reagent volume for one rinsing cycle.

Cycles

Number of rinsing cycles to be carried out.

Dosing

rate

The maximum value for the dosing speed depends on the
cylinder size of the buret used.
Buret cylinder

Max. speed

1 mL

3 mL/min

2 mL

6 mL/min

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

For smaller buret cylinders a value of up to 150 mL/min

can also be entered, but when the function is carried out

the dosing speed will automatically be reduced to the