beautypg.com

9 cal calibration of ph sensors – Metrohm 726 Titroprocessor User Manual

Page 122

background image

7.9 CAL

Calibration of pH sensors

726 Titroprocessor, Instructions for use

116

116

7.9 CAL

Calibration of pH sensors

Application:
The calibration of a pH sensor is indispensable not only for pH measure-

ments; it is also essential for SET titrations and for the correct determination

of pK values and fixed endpoints.

In contrast to manual sensor calibration, the CAL mode allows the calibra-

tion parameters and individual buffer series to be set down in a method. The

calibration data, like all determination data, can be stored and are available

for subsequent data output; however, they cannot be reactivated.

Calibration principle

During the pH calibration an isotherm (constant temperature) of the pH

value is recorded as a function of the electrode system voltage (electrode

system = sensor + electrode cable + connection socket + measurement

interface) and used for the conversion of the measured voltages into pH

values. This means that all the buffer solutions used for the calibration must

be at the same temperature, even with automatic temperature correction

using a temperature sensor. If possible the calibration should be carried out

at the same temperature as that at which the later measurements will be

carried out.

pH

14

7

0

100

200

300

400

500

-500

-400

-300

-200

-100

U[mV]

d U

d pH

*

pH

as

Slope =

dU/dpH

U

N

(T)

Temperature 25°C

Without calibration the relative slope of the electrode system will be 1

and pH

as

will be set at 7.

Depending on the requirements for accuracy the calibration can be carried

out with 1, 2 or up to 9 buffer solutions.