88. . Command Reference

Command Reference

726 Titroprocessor, Instructions for use

129

129

Assignment parameters of the DET modes:

comment

Comment on titration mode

(52 characters)

The comment on the mode serves to identify the specific

titration in a determination. This is particularly useful for

more complex methods with several titration or measuring

modes.

Reagent

Name of the reagent

(13 characters, selectable)

If a name is entered here, the automatic reagent control

can be used. The field will have a list of items for selection if

a reagent name has already been assigned under 'De-
vices & Manual Control

'.

When the method is started, the entry under 'Devices &
Manual Control

' ( key) for the cur-

rent occupancy of the addressed dosing drive is checked.

If non-consistency is found an error message will be indi-

cated.

Leaving the 'Reagent' field empty (or on the Devices &
... page ) prevents an automatic reagent control.

Sensor

Name of the sensor to be used

(13 characters)

If a name is entered here, the automatic sensor control can

be used.
When the method is started, the entry under 'Devices &
Manual Control'

( key) for the

connected sensor is checked. If non-consistency is found

an error message will be indicated.

Leaving the 'Sensor' field empty (or on the 'Devices & ...
page ) prevents an automatic sensor control.

Concentr.

Concentration of the reagent

(numerical | 5 characters)

Two fields for entering the concentration of the reagent so-

lution and the unit of concentration. These fields are also

(see above) used for the reagent control and, when the

method is started, are carefully compared with the entry in

the 'Devices & Manual Control' dialogue window.
The reagent concentration can be used for calculations. It

is filed in the C system variable.