22. . Installation

Installation

726 Titroprocessor, Instructions for use

23

23

2.6.3 External bus

The external bus allows external instruments to be controlled by the 726 Ti-

troprocessor. Sample changers and dosing devices such as the 685 Dosi-

mat or the 700 Dosino can communicate bi-directionally with the Titroproc-

essor (=bus master).

External devices must be connected with EBus cables according to the fol-

lowing scheme:

Bus master

External device

External device

External bus

Each external instrument is identified by an EBus address. When the Titro-

processor is switched on they are automatically identified. An EBus address

must be set for each external device (1…9, A…F; 0 stands for the bus

master).

Address selector

The corresponding EBus address can be set with

a screwdriver. The EBus address selector is usu-

ally located on the rear panel of the external de-

vice.

External device

EBus

address

Corresponding

peripherals

Sample changer 1

1

1…2 lifts, 0…4 pumps,

1…4 stirrers

Sample changer 2

2

1…2 lifts, 0…4 pumps,

1…4 stirrers

Dosimat interface 1

1

1…4 Dosimats (B1…B4)

Dosimat interface 2

2

1…4 Dosimats (C1…C4)

Devices such as sample changers which consume a lot of current should

be connected in the first position of the EBus chain. This is particularly im-

portant if the relevant sample changer is equipped with more than one

pump. The 729 Dosimat interfaces should be connected to the 717 Sample

changer.

The 726 Titroprocessor and the Dosimat interfaces must be connected to

the same earth. A line distributor should be used to supply all devices on

the external bus.

The 726 Titroprocessor must be switched off before a pe-
ripheral device is connected.