88. . Command Reference

Command Reference

726 Titroprocessor, Instructions for use

135

135

It is preferable to select a relative stop volume when the

primary objective is to limit the consumption of reagent.

Fill. rate

Buret fill rate

0.01 mL/min…150 mL/min | max.

The buret is automatically filled when the titration ends. The

maximum fill rate will depend on the size of the buret cylin-

der used.

Buret cylinder

max. fill rate

1 mL

3 mL/min

2 mL

6 mL/min

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

A value up to 150 mL/min can be entered even for small

buret cylinders. When the method is executed, however,

the dosing rate will automatically be reduced to the largest

possible value. Select 'max' whenever possible, except with

viscous reagent solutions which could cause underpres-

sure in the cylinder if the filling rate is too fast (formation of

bubbles!).

Meas.val

Measured value as a discontinuation criteria

with DET_PH:

pH -20 … 20 | off

with DET_U, DET_IPOL:

-2000 … 2000 mV | off

with DET_UPOL:

-200 … 200 µA | off

The measured value at which the titration is discontinued.

EP#

Number of equivalence points

0…9 | off

The titration is discontinued as soon as the pre-set number

of equivalence points is recognized. In principle, only 9

equivalence points can be evaluated. In the setting

'EP# 0' the titration is discontinued after the start condi-
tions are executed (initial dosing). 'off' means, the num-
ber of equivalence points found is not taken into account

for the discontinuation of a titration.

Stop time

Absolute titration duration

1…9999 s | off

The titration is discontinued on expiry of the pre-set stop

time.