22. . Installation
Installation
726 Titroprocessor, Instructions for use
17
17
2.3.2 Mains cable and mains connection
The instrument is supplied with one of three types of cable:
•
6.2122.020 with plug SEV 12 (Switzerland,
…
)
•
6.2122.040 with plug CEE(7), VII (Germany,
…
)
•
6.2133.070 with plug NEMA 5-15 (USA,
…
)
which has three leads and is fitted with a plug with an earthing pin. If a dif-
ferent plug has to be fitted then the yellow/green lead (IEC standard) must
be connected to the earth (protection class 1). If no earthed socket is avail-
able then a proper earth connection must be made via the earthing socket
of the instrument.
Any break in the earthing lead, whether inside or outside the in-
strument, can cause the instrument to become a hazard!
Insert the mains cable into the mains connector of the 726 Titroprocessor.
2.3.3 Setting up the thermal printer
Insert paper (only for the built-in printer)
The thermal printer of the Titroprocessors (models 2.726.0010 and
2.726.0110) is supplied without inserted paper. The thermal paper
6.2237.040 included with the accessories (roll approx. 40 m
≅
130 A4
pages) is inserted into the printer as follows:
Preparations
•
Switch on 726 Titroprocessor.
•
Lift up the printer cover.
•
Make sure that the roll of thermal paper
6.2237.040 has a straight edge; if necessary,
carefully tear off the paper at the first perfora-
tion.
Insertion
•
Insert paper spool 6.2241.020 in thermal paper roll 6.2237.040.
•
Place the roll of paper with the paper spool in the two recesses at the
sides of the printer so that the paper leaves the roll from the rear and the
file hole margin is on the left-hand side when seen from the front.
•
Carefully insert the straight-edged paper manually into the paper guide
slot to the limit stop and hold it there.
Paper transport
•
Press the two keys
be fed in automatically and transported to the correct position.
•
Close the cover of the paper storage compartment.