beautypg.com

Metrohm 726 Titroprocessor User Manual

Page 261

background image

12

12. . Troubleshooting

Troubleshooting

726 Titroprocessor, Instructions for use

255

255

TITRATION 8:

titration start while not cond.ready

A titration has been started although the previous conditioning mode was

not finished correctly. The started titration may produce an incorrect result.

TITRATION 9:

no current sample size

No sample size has been entered. Enter a sample size on the results page

(Softkey [Sample data]) and allow the determination to be recalculated

(Softkeys [ >> >>] and [Recalc one]).

TITRATION 10:

dosing drive not responding

The connection to the selected dosing drive has been interrupted during the

titration. Stop the determination, switch off the Titroprocessor and reconnect

the dosing drive. Switch the Titroprocessor on again.

TITRATION 11:

ADD_R volume too high, corrected to 1000 mL

The calculation of the dosing volume for the ADD_R command from the

factor and sample size has resulted in too large a value. The largest possi-

ble volume is 1000 mL. If the error message is confirmed with the

dosing volume will be corrected to 1000 mL.

TUBINGS

TUBINGS 1:

same name for type no. "X" and no. "Y"

At least two tubing sets have been given the same type name. The names

must be unambiguous.

TUBINGS 2:

type no. "X" is not complete

The parameters for the definition of buret unit no. X have not been com-

pletely defined.