background image

8. Command reference

8. Command reference

726 Titroprocessor, Instructions for use

207

207

This also applies when the sample silo is being used. In

this case the entries in the input fields can be edited un-

der [Sample silo] / [>> >>] [Sample
idents]

.

none

no sample ident queries

1

only query Ident1

2

only query Ident2

3

only query Ident3

1+2

query Ident1 and Ident2

all

query Ident1…3

The input of a sample remark is always possible in the

ASK window.

Size

Query mode for sample weight

no | yes | confirm.

A sample weight query can be made in different ways.

no

no sample weight query

yes

after sample weight input the ASK
window is closed automatically

confirm.

Sample weight input possible, the
ASK window must be closed with

.

HOLD

Interrupt method run

The method run can be interrupted with the HOLD command. From the

Hold condition the method can be restarted with the key

or . In the Hold condition the yellow keyboard keys can be used

(e.g. , ). In this way it is possible to operate

the different peripheral devices in the Hold condition.

Parameters

Display text of HOLD window

50 characters

A text can be entered for display in the Hold window. Ex-

ample:

ЪДДДДДДДДДДДДДДДДДДДДДД

Message

ДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ Hold: ³

³ ------- Add sample now! ------------------------ ³

³ Press ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙ

WAIT

Waiting time in method run

The method run can be interrupted for a preset time (in seconds) with the

WAIT command. The duration of the interruption is defined with the line time

of the command line. If no line time is given then the method run will remain

interrupted until the WAIT display window is closed with the key.

This is also possible if a defined waiting time has to be prematurely termi-

nated.