Metrohm 726 Titroprocessor User Manual
Page 216
8.14 Auxiliary commands
726 Titroprocessor, Instructions for use
210
210
Command Parameters Parameters
ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД
SHOW
Page main
Page
Show dialog page
curve | result |stat.| main
Depending on the selected dialog page the switch is ei-
ther made immediately or at the end of the determination.
curve
immediate switch to live curve
display
result
switches to results display at end of
determination
stat.
switches to statistics results display at
end of sample series
main
display immediate switches to
main page
NOOP
No operation, place holder command
The NOOP command can be used to arrange the command blocks. It does
not trigger any actions in the method run.
BEEP
Acoustic signal
The BEEP command is used to produce an acoustic signal.
END
End marker
Each method must contain an END command. This is used to mark the end
of the method. Normally the END command is set in the last method line.
For test purposes in a complex method the insertion of further END com-
mands may be useful, e.g. in a CASE sequence. The method will then be
processed until the first END marker is reached.
STIR_I
Control of internal stirrer connections
The stirrer connections of measuring groups A and B can be switched di-
rectly in a method. Stirrer models 728 (magnetic stirrer), 722 (rod stirrer) or
the 703 or 727 Titration stands (with magnetic stirrer) are suitable for use on
the internal connections of the measuring interfaces.
Command Parameters Parameters
ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД
STIR_I
Meas. interface A State on
After switch-on the stirrer connections are also switched on. The stirrers
connected to the measuring groups can also be operated manually; see