8.7 KFC

726 Titroprocessor, Instructions for use

170

170

A value up to 150 mL/min can also be entered for smaller

buret cylinders; however, when the method is carried out

the dosing speed will automatically be reduced to the larg-

est possible value. For KF reagents (viscous liquids) no

filling rate greater than 30 mL/min should be selected as

otherwise too rapid filling may cause a partial vacuum in

the cylinder (bubble formation!).

Temperature

Temperature setting or measuring mode

-170…25.0…500°C | auto | initial

The temperature can be automatically determined during a

titration with the aid of a Pt100/Pt1000 sensor or a com-

bined electrode with built-in temperature sensor. It is pos-

sible to record the temperature throughout a titration and to

output it as a temperature curve.

numerical

manual input

auto

automatic measurement during the
whole titration (for temperature curve)

initial

one-off measurement at the start of the
titration mode

The temperature can be included in formulas for calcula-

tions (see page 118ff), e.g. for volume corrections. The dif-

ferent system variables are listed on page 148 and 128.

Please note: Start temperature ST and closing tempera-

ture CT are only available with auto.

U(pol)

Polarization voltage (with KFC_UPOL*)

-1270…400…1270 mV

The polarization voltage is the voltage which is applied to a

polarized electrode during amperometric titration. The re-

sultant measuring quantity is I

pol

/µA.

I(pol)

Polarization current (with KFC_IPOL*)

-127…25…127 µA

The polarization current is the current which is applied to a

polarized electrode during voltametric titration. The resul-

tant measuring quantity is U

pol

/mV.