8.11 Dosing commands

726 Titroprocessor, Instructions for use

192

192

Dos. rate

Addition speed

0.01 mL/min…150 mL/min | max.

The maximum dosing speed depends on the cylinder size

of the buret used.

Buret cylinder

max. speed

1 mL

3 mL/min

2 mL

6 mL/min

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

A value up to 150 mL/min can also be entered for smaller

buret cylinders; however, when the method is carried out

the dosing speed will automatically be reduced to the

largest possible value. Whenever possible 'max' should

be selected, except for viscous reagent solutions.

Volume

Absolute volume

0…999.99 mL

When entering the absolute reagent volume please take

the capacity of the target vessel into account.

Fill. rate

Buret filling speed

0.01 mL/min…150 mL/min | max.

The filling rate set here applies to both a concluding auto-

matic filling after a titration and possible refilling during a ti-

tration. The maximum value for the filling rate depends on

the cylinder size of the buret used.

Buret cylinder

max. speed

1 mL

3 mL/min

2 mL

6 mL/min

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

A value up to 150 mL/min can also be entered for smaller

buret cylinders; however, when the method is carried out

the dosing speed will automatically be reduced to the

largest possible value. Whenever possible 'max' should

be selected, except for viscous reagent solutions for

which too rapid filling can cause a partial vacuum in the

cylinder (bubble formation!).