22. . Installation

Installation

726 Titroprocessor, Instructions for use

27

27

The data transmitted from the balance are directly interpreted by the 726 Ti-

troprocessor within the permitted input range for the sample weight includ-

ing sign and decimal point. The units given in the following table are also

accepted:

Gram

g

Milligram

mg

Kilogram

kg

Pieces

pcs

Pieces

PC

Carat

ct

Pound

lb

Ounce

oz

Troy ounce

ozt

Grain

GN

Pennyweight

dwt

Any control characters which the balance may transmit will cause the Titro-

processor to produce an error message. Switch off all your balance's auto-

matic status messages when it is connected to the 726 Titroprocessor.

With the aid of a special input unit supplied by the balance manufacturer it is

possible to enter the sample identification and method from the balance as

well as the sample weight. This is done by preselecting the addresses of the

identification and method on the input unit.

Balance

Balance

Method

Method

Ident1

Ident1

Ident2

Ident2

Ident3

Ident3

Sartorius

METH or 27

ID.1 or 26

ID.2 or 24

C-20 or 23

Mettler (AT)

D (Mthd)

C (ID

#

1)

B (ID

#

2)

A (c20)

Units which are not listed and

which are transmitted directly

after the weight cannot be in-

terpreted and will therefore be

rejected.