Metrohm 726 Titroprocessor User Manual
Page 271
13. Appendix
726 Titroprocessor, Instructions for use
265
265
13.1.7 To check Remote interface
With this diagnostic step the interface Remote can be checked for its functionality. For this
test a special test plug with article number 3.496.8510 is required.
1.
Prepare instrument for diagnostic test (see chap. 13.1.2).
2.
Without switching off the instrument plug in test plug 3.496.8510 to socket
"Remote".
3.
Ä
The test runs automatically. If a defect is found, an error message appears on the display, otherwise the previous
screen page appears.
4.
Without switching off the instrument remove test plug 3.496.8510.
13.1.8 To check RS232 interface
With this diagnostic step the two interfaces RS232 can be checked for their functionality.
For this test a special test plug with article number 3.496.8480 is required.
1.
Prepare instrument for diagnostic test (see chap. 13.1.2).
2.
3.
Without switching off the instrument plug in test plug 3.496.8480 to socket "RS232
Interface 1".
4.
Ä
The test runs automatically. If a defect is found, an error message appears on the display, otherwise the previous
screen page appears.
5.
6.
Without switching off the instrument plug in test plug 3.496.8480 to socket "RS232
Interface 2".
7.
Ä
The test runs automatically. If a defect is found, an error message appears on the display, otherwise the previous
screen page appears.
8.
Remove test plug 3.496.8480.
9.
The test is quit by pressing