Metrohm 726 Titroprocessor User Manual
Page 41
3. Configuration
726 Titroprocessor, Instructions for use
35
35
Ident1,2,3+Meth
sample identification+
method name
Ident1+Remark
sample ident.+ remark
Ident1,2+Remark
etc.
Ident1,2,3+Remark
…
Ident1+Meth+Remark
…
Ident1,2+Meth+Remark
…
The scheme for the automatic creation of file
comments corresponds to that for automatic file
name creation, see above.
Power on
PREP warning
If this warning is switched on this means that
immediately after the Titroprocessor is switched on a
display window appears with a reminder that a
dosing device must be prepared before the start of a
sample series , i.e. the buret must be rinsed with
reagent.
Remote
terminal mode
This mode can be switched on by the remote control
of the Titroprocessor 726 via the RS232 interface. It has the effect of
sending the screen contents with all character attributes as ANSI control
sequences via the RS232 interface as soon as an alteration to the screen
contents occurs.
File transfer
protocol
For data storage of determination or other files on a
personal computer two different data transfer
protocols are available. If Metrodata 726 backup
software is used then 'Metrohm (Bkup)' should
be selected, otherwise select 'Kermit'. This latter
requires that a PC terminal program is used which
supports this data transfer protocol.
Softkeys
[Device
header]
opens the dialog window for the entry of an
instrument-specific, four-line header, which will be
printed out before the report of a determination.
[File]
opens the file dialog window for storing and loading
configuration files. It is recommended that all the
instrument configuration settings are stored on a
data card in this way. In a configuration file all the
subordinate areas on the configuration page are
stored in addition to the common variables. The rack
tables of special racks are not stored and, if
necessary, must be separately loaded with the help
of rack files, see page 40.