beautypg.com

5 sec endpoint conditioning, 1 control of the conditioning mode – Metrohm 726 Titroprocessor User Manual

Page 114

background image

7.5 SEC

Endpoint conditioning

726 Titroprocessor, Instructions for use

108

108

Observation

Possible causes and remedies

"Overshoots". Titration

is not settled,

i.e. single pulses

are not added at

the end.

Reduce 'Max. Rate'.

Set a larger control range.

Set a much smaller 'Min. Rate'.

Optimize the arrangement of the electrode and buret tip and improve the

stirring. This is particularly important for very rapid titration reactions and

for steep curves.

Titration time

is too long.

Set higher 'Min. Rate'.

Set higher 'Max. Rate'.

Reduce 'Control range'.

Titration results

scattered too

much.

Set lower 'Min. Rate'.

7.5 SEC

Endpoint conditioning

Application:

SEC modes are independent titration modes which can be used in the

preparation of a SET mode. They are suitable for rapidly bringing a titration

sample to a certain measuring value (endpoint).

Their typical feature is controlled reagent dosing depending on the control

difference (actual value – endpoint) with the aim of reaching the preset value

as quickly and accurately as possible. Overshooting is virtually eliminated.

No provision is made for the representation of curves or carrying out calcu-

lations.

Reagent addition and switch-off criteria correspond to those of the SET

modes.

7.5.1 Control of the conditioning mode

Conditioning is normally carried out in the background, i.e. as soon as a

conditioning mode has been started the 726 Titroprocessor can carry out

other commands (i. e. ASK, ADD_A etc.). A conditioning process also re-

mains active even when a determination has been finished, provided that it

has not been explicitly terminated; see below.

A conditioning mode carried out in the background is controlled by the

WAIT_C command (see page 208f). This checks whether the conditioned

state has been achieved and, if necessary, switches the conditioning mode

off.