7.10 Calculations / Formulas

726 Titroprocessor, Instructions for use

120

120

stat.

The result enters the statistics.

live

The result is shown in the trace win-
dow during the determination and

recalculated each time a new entry

is made in the measuring point list.

Please note that in titrations the live

value may not be quite as up-to-

date as the actual value.

Softkeys

[Copy line]

copies the contents of the selected line into a local
buffer memory.

[Cut line]

copies the contents of the selected line into a local
buffer memory and deletes the selected line.

[Paste line]

inserts a new line before the selected line and enters
the contents of the local buffer memory in it.

[Insert line]

inserts a new empty line before the selected line.

[Delete line]

deletes the selected line.

7.10.1 System variables

The following system variables can be used in formulas.

(xx=any 2 numbers or letters, #=any number)

Rxx

Result

RT1

Result for validation

RT2

Theoretical result for validation

RE

Result for calculating the system error in validation

Xxx

Intermediate result

C##

Common variable (of the actual titration or measurement)

EP#

Endpoint volume in mL

EM#

Endpoint measurement (unit depends on mode)

EF#

Endpoint flag; endpoint found yes/no (1/0)

F#

Fixed endpoint volume in mL

FM#

Fixed endpoint measurement (unit depends on mode)

FF#

Fixed endpoint flag; fixed endpoint reached yes/no (1/0)

MNRxx

Mean value of results Rxx in statistics memory

MNXxx

Mean value of intermediate results Xxx in statistics memory

MNC##

Mean value of common variables C##

ID1

Sample identification 1

ID2

Sample identification 2

ID3

Sample identification 3

SS

Sample size (value without units)

DD

Determination duration in s, time for whole determination

MD

Measuring duration in s; time for actual titration or measure-

ment

SD

Starting duration in s; length of start phase of a mode

IT

Initial temperature in °C

ST

Starting temperature in °C (after start phase), only for auto-

matic measurements.

CT

Closing temperature in °C, only for automatic measurements