8.4 SET

726 Titroprocessor, Instructions for use

152

152

controlled addition will be carried out throughout the

whole titration.

U[mV]

V[mL]

EP

Dynamics

Min. rate

Smallest possible titration speed

0.01…25…999.9 µL/min | min.

In order to avoid excessively slow titration speeds (and

thus "never-ending" waiting times) during dynamically

controlled titrations the smallest possible titration speed

can be set. This parameter should only be altered with

caution, as too rapid reagent addition can negatively in-

fluence the accuracy of the results.

Max. rate

Largest possible titration speed

0.01…10.00…150 mL/min | max

Setting the max. rate has a direct effect on the dosing

speed of the titration control. The maximal value de-

pends on the cylinder size of the buret.

Buret cylinder

max. speed

1 mL

3 mL/min

2 mL

6 mL/min

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

A value up to 150 mL/min can also be entered for

smaller buret cylinders; however, when the method is

carried out the dosing speed will automatically be re-

duced to the largest possible value. Whenever possible

'max' should be selected, except for viscous reagent

solutions.