8.2 DET

726 Titroprocessor, Instructions for use

142

142

Quantity X_Axis

Display unit on the X-axis

Volume | Time

Titration curves can be displayed against volume or

time. The time axis can only be displayed if the time re-

cording function (see page 134) is switched on.

Scale

Type of axis scale

auto | fixed

The two axes of the curve display can be automatically

scaled by the Titroprocessor. If a fixed scale is required,

the operator must indicate all the scaling parameters

(see below).

begin … end

Limit values for a fixed scale

Y-axis:

pH -20…20 | -2000…2000 mV

-200…200 µA | ERC 0…200

X-axis:

0…999,99 mL | 0…999 999 s

The limit values of the scale can be chosen as required

within the limits defined by the measurable variables. If

you enter fixed limit values, ensure that the scale func-

tion is also set to 'fixed'.

Division

Intervals between scale marks on the X-axis

0,01…1…99,9 mL/cm

0,01…10…99,9 s/cm

The scale division on the X-axis only applies for printing

out titration or measuring curves.

Grid

Titration curve grid

on | off

The titration curve grid can only be viewed when the

curve is printed out. It cannot be viewed on the screen.