88. . Command Reference

Command Reference

726 Titroprocessor, Instructions for use

131

131

The various conditions governing the discontinuation of initial dosing can be

combined at will.

N.B.: Initial dosing is discontinued on attainment of any of the start condi-

tions, be it the defined absolute/relative start volume, the defined measured

value or the signal slope. The actual titration will then start immediately after

the pre-set pause.

predef. start volume

meas. value

(reached)

Slope

Vol [mL]

U [mV]

actual initial dosing

The volume actually dosed is filed in the SV system variable and can be

called up for calculations.

Volume

Definition of the start volume

absolute | relative | off

The defined start volume can be entered as an absolute

volume, or as a relative volume as a function of the sample

size (recommended). Various other fields* will need to be

edited depending on what is entered in this field

(see be-

low).

Size

**

Absolute volume (with abs. start volume)

0…999.99 mL

If an absolute start volume is entered, the start volume

should be kept small enough to ensure that sufficient vol-

ume remains up to the equivalence point for the dynami-

cally controlled titration.

Factor

**

Calculation factor (with rel. start volume)

–99999…0…99999

The actual volume of the start volume to be added will be

calculated as follows:

Volume in mL = sample size

∗∗ factor

.

The maximum start volume that can be added is 1000 mL.

If the calculation results in a higher value, the start volume

is automatically reduced to 1000 mL.