beautypg.com

Metrohm 726 Titroprocessor User Manual

Page 186

background image

8.9 CAL

726 Titroprocessor, Instructions for use

180

180

The temperature can be included in formulas for calcu-

lations (see page 118ff). The different system variables

are listed on page 178 and 128.

8.9.2 Specific CAL* mode parameters [ * Param ]

The buffer tables for the relevant CAL* mode can be found under the
[* Param ]

softkey.

ЪДДД

CAL* Parameters

ДДДДДДД¿

³ ³

³ Buffer Type Value [pH] ³

³ ДДДДДД ДДДДДДД ДДДДДДДДД ³

³ 1 Metrohm 7.000 ³

³ 2 Metrohm 4.000 ³

³ 3 Metrohm ³

³ 4 Metrohm ³

³ 5 Metrohm ³

³ 6 Metrohm ³

³ 7 Metrohm ³

³ 8 Metrohm ³

³ 9 Metrohm ³

³ ³

АДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙ

Up to nine different buffer solutions of various types or manufacturers can

be entered here.

Type

Buffer solution type or manufacturer

Metrohm | NIST | DIN | Fisher | Novartis | Ingold |

Merck | Beckman | Radiom. | none | custom

As the buffer solutions can have very different composi-

tions, it is necessary to enter the correct buffer type so

that the temperature correction function can be correctly

applied during the calibration.

If your type of buffer is not given in the list you should

select 'none'.
During the calibration the determined values are com-

pared with the buffer solution information (theoretical

value, temperature for automatic temperature meas-

urements) and subjected to a plausibility test, see page

117. If this leads to interfering error messages with 'ex-

otic' buffers or 'extraordinary' sensors, then select the

buffer type 'custom' for user-defined buffer types. No
plausibility test is carried out in this setting.

Value [pH]

Nominal pH of the buffer

pH -20…20

The buffer solutions offered by the above-mentioned

manufacturers are stored in the Titroprocessor with their

nominal values. The entry field has a selection list in

which various buffer values depending on the given

buffer type or manufacturer can be selected. It is also