4. Manual operation

Manual operation

726 Titroprocessor, Instructions for use

59

59

Measuring parameters

Signal drift

Limits for signal drift

0.5…2…999 mV/min | off

The measurement is only accepted if the measurement

change of the sensor falls below the preset value for the

signal drift. If this is delayed, the measurement will

nevertheless be accepted when the preset equilibration

time (see below) has elapsed.

This signal drift control can be switched off ('off').
However, this cannot be recommended for correct

calibration.

Equilibr.time

Max. equilibration time before measurement
acceptance

0.1…9999 s | auto | off

The whole equilibration time for measurement

acceptance is only effective when, with drift control

switched on (see signal drift above), the preset value for

signal drift is not undercut. Otherwise the measurement

is accepted as soon as the preset signal drift is

achieved.

The whole equilibration time will also be allowed to

elapse if the drift control is switched off.

The equilibration time can be switched off. This means

that only the signal drift is decisive for measurement

acceptance.

In the setting ‘Equilibr.time auto' the
Titroprocessor will calculate an equilibration time suited

to the preset value for the signal drift according to the

following equation:

Equilibr time in s

drift

mV

.

(

)

/ (

/ min)

,

=

+

+

150

0 01

5

For the standard setting 2 mV this results in an

equilibration time of 110 s.

Temperature

Temperature setting or measuring mode

-170…25.0…500°C | auto | off

The temperature can be determined during a

measurement with the help of a Pt100/Pt1000 sensor or

a combined electrode with built-in temperature sensor.

It is required for automatic pH correction.

numeric

manual input

auto

automatic measurement during the

calibration

off

no temperature correction