beautypg.com

Metrohm 726 Titroprocessor User Manual

Page 69

background image

4. Manual operation

Manual operation

726 Titroprocessor, Instructions for use

63

63

Softkeys

[Reset

changer]

initializes the sample changer. When a sample rack has
been exchanged it is necessary to initialize the sample

changer. In this process all lifts are moved to the rest

position, the sample turntable is rotated to the initial

position and the magnet code is read to identify the new

rack.

[Remote

lines]

opens the dialog window of the remote-lines of the
selected sample changer. The conditions of the logical

input lines can be seen and the conditions of the output

lines can be set.

ЪДДД

Remote lines: 1

ДДДДДД¿

³ ³

³ I/O lines 32109876543210 ³

³ ДДДДДДДДДДДДДДДДДДДДДДДД ³

³ Input 00000001 ³

³ Output

**********0001

³

³ ³

АДДДДДДДДДДДДДДДДДДД QUIT ÄÙ

The output lines can be set individually (1) or reset (0).

Lines which are not to be altered can be given an '*'.

Further details can be found on pages 200f and 29f.

[Stirrer

on/off]

switches the selected stirrer on or off.

[Pump

on/off]

switches the selected pump on or off.

[Lift up]

moves the lift of the selected tower upwards. This is only
possible if the sample rack is aligned to the tower. With

sample racks which are not expressly designed for two-

tower operation e.g. 14-place or 16-place racks, the rack

should first be rotated to the left or right by one position,

see softkeys below.

[Lift

down]

moves the lift of the selected tower downwards. Please
note the remarks about the previous softkey.

[Shift

left]

rotates the sample rack of the selected changer
counterclockwise by one position. The lifts of the changer

must be in the shift position or a higher position.

[Shift

right]

rotates the sample rack of the selected changer
clockwise by one position. The lifts of the changer must

be in the shift position or a higher position.

[Shift

to pos.]

opens the dialog window for targeted positioning of
sample racks, see below.