4.4 Basic settings

732 IC Detector / 733 IC Separation Center

91

>CONFIG/print meas.value
time interval 1.0 s

0.4

99999 s

Time interval for measured value printout

This inquiry appears only with "

print crit. = time

".

Time interval in seconds between printout of the indi-

vidual measured values.

>CONFIG/print meas.value
stop time off min

1

999 min, off

Stop time for measured value printout

This inquiry appears only with "

print crit. = time

".

Time to stop of the measured value printout.

off

No time restriction

(press ).

>CONFIG/print meas.value
date&time: off

on, off

Printout of date and time

This inquiry appears only with "

print crit. = immed.

" or "

print

crit. = time

".

on

Date and time are printed out for every

measured value.

off

Date and time are not printed out.

>CONFIG
>CONFIG/auxiliaries

General instrument settings

>CONFIG/auxiliaries
run number 0

0

999, off

Run number
The run number is incremented by +1 for each new

determination as follows:

without program Increase with every "INJECT A".
with program

Increase with every new start and

with every "

return

" in cycle pro-

grams.

The parameter value has the following meaning:

0

999

Start point for numbering.

off

No numbering

(press ).

>CONFIG/auxiliaries
number of cycles 1

1

999

Number of cycles
This parameter is identical to the parameter of the same

name under the key (see section 4.7.1)

and determines how many times cycle programs should

be run. As access to the key is blocked

while a program is running, the number of cycles can

subsequently be changed here.