Metrohm 732 IC Detector User Manual

Page 46

background image

2.8 Separating columns and suppressor module

732 IC Detector / 733 IC Separation Center

39

Attach inlet capillary 90

90 using a 6.2744.010 Compression

fitting to the connector 55

55 of the filter unit PEEK (see section

2.6.3).

Attach a piece of the 6.1803.020 PTFE tubing (from 752

accessories) cut to the required length using a 6.2744.010

Compression fitting to the connector 53

53 with filter at the other

end of the Filter unit PEEK.

Attach the other end of the PTFE tubing using a 6.2744.010

Compression fitting to the 6.2744.030 Coupling (from 752 ac-

cessories) and mount it on the outlet end of the first length of

pump tubing.

Mount a 6.2744.030 Coupling to the inlet end of the first

length of pump tubing. Attach a piece of the 6.1803.020 PTFE

tubing (aspirating tubing, from 752 accessories) cut to the

required length using a 6.2744.010 Compression fitting to the

other end of this coupling.

Immerse the other end of the aspirating tubing in a vessel

containing regeneration solution (normally 20 mmol/L H

2

SO

4

)

and fix in place.

Pull outlet capillary 91

91 marked with "Waste" of the suppressor

module through the rear panel opening 43

43, lead it to a suffi-

ciently large waste container and fix it in place.

9 Suppressor connection 3: H

2

O

Loosen rotary nipple screwed onto the interior side of con-

nection 28

28. Pull inlet capillary 93

93 marked with "H

2

O" (see

Fig. 17 and Fig. 18) by hand out of the opening of connection

28

28 as far as required. Retighten nipple on the interior side of

connection 28

28 to fix inlet capillary 93

93.

Attach inlet capillary 93

93 using a 6.2744.010 Compression

fitting to the connector 55

55 of the filter unit PEEK (see section

2.6.3).

Attach a piece of the 6.1803.020 PTFE tubing (from 752

accessories) cut to the required length using a 6.2744.010

Compression fitting to the connector 53

53 with filter at the other

end of the Filter unit PEEK.

Attach the other end of the PTFE tubing using a 6.2744.010

Compression fitting to the 6.2744.030 Coupling (from 752 ac-

cessories) and mount it on the outlet end of the second

length of pump tubing.

Mount a 6.2744.030 Coupling to the inlet end of the second

length of pump tubing. Attach a piece of the 6.1803.020 PTFE

tubing (aspirating tubing, from 752 accessories) cut to the

required length using a 6.2744.010 Compression fitting to the

other end of this coupling.

Immerse the other end of the aspirating tubing in a vessel

containing rinsing solution (normally dist. H

2

O) and fix in

place.

Pull outlet capillary 92

92 marked with "Waste" of the suppressor

module through the rear panel opening 43

43, lead it to a suffi-

ciently large waste container and fix it in place.