5 Notes – Maintenance – Faults

732 IC Detector / 733 IC Separation Center

150

5.4.2

Prepare instruments

Power off.

Disconnect all external connections at the 732 IC Detector

(except mains cable).

Power on and immediately press and hold <9> key until

initialisation EEPROM?
ENTER>yes '9'>diagnosis

Press <9> key to open the following diagnostic menu:

diagnosis
>diag/EEPROM initial.

Initialize data memory

1)

sectn 5.4.9

diagnosis
>diag/run mode

Parameters for sequence control of
“internal test“ and “external test“

2)

diagnosis
>diag/RAM test

Check working memory (RAM)

3)

sectn 5.4.3

diagnosis
>diag/keyboard test

Check keypad

3)

sectn 5.4.4

diagnosis
>diag/display test

Check display

3)

sectn 5.4.5

diagnosis
>diag/RS test

Check RS232 interfaces

3)

sectn 5.4.6

diagnosis
>diag/IO test

Check remote interfaces

3)

sectn 5.4.7

diagnosis
>diag/internal test

Internal hardware test

3

sectn 5.4.8

diagnosis
>diag/external test

External hardware test

2)

diagnosis
>diag/instr.adjustment

Instrument adjustment

2)

diagnosis
>diag/instrument number

Reading and writing the instrument
number

2)

diagnosis
>diag/power on reset

Power on reset

2)

Pressing the <9> key again selects the submenus in turn. The individual inquiries

of a submenu are accessed using the key, exit is with the key.

1)

Initialization:

This diagnostic step is intended for the initialization of the data memory.

2)

Care:

These diagnostic steps are reserved for the service engineer and are not described.

3)

Test functions:

These diagnostic steps are test functions which the customer can perform.