5 Notes – Maintenance – Faults
732 IC Detector / 733 IC Separation Center
142
3
Clean input and output leads
•
Connect each of the 6 capillary tubings attached to connec-
tion piece 98
98 to the pump one after another, and pump
through ultrapure water.
•
Check whether solution emerges from connection piece 98
98.
If one of the input or output leads remains blocked, replace
the connection piece (order number 6.2832.010).
4 Clean suppressor rotor
•
Clean the sealing surface of suppressor rotor 99
99 using a lint-
free cloth and ethanol.
5 Insert suppressor rotor
•
Insert suppressor rotor 99
99 in suppressor holder 100
100 in such a
way that the tubing connections at the rear of the rotor fit in
the corresponding openings inside the rotor, and that one of
the three holes in the rotor can be seen from below in one of
the openings of the holder.
•
If the rotor has been inserted correctly, its sealing surface will
be about 4 mm inside the holder. If this is not the case, bring
the rotor into the correct position from below with the aid of a
sharp object (e.g. a screwdriver).
6 Clean connection piece
•
Clean the sealing surface of connection piece 98
98 using a lint-
free cloth and ethanol.
100
100
99
99
98
98
97
97
Fig. 24: Assembling the suppressor
97
97
Screw nut
99
99
Suppressor rotor (6.2832.000)
98
98
Connection piece
(6.2832.010)
with input and output leads
100
100
Suppressor holder