Metrohm 732 IC Detector User Manual
Page 198

7.1 Technical data
732 IC Detector / 733 IC Separation Center
191
Interfaces
RS232 interfaces
Specifications, see section 6.1
Remote interfaces
Specifications, see section 6.2
Analog output
Specifications, see section 6.3
Safety specifications
Construction/testing
According to IEC 1010 / EN 61010 / UL 3101-1,
protection class 1, degree of protection IP40
Safety directions
The Instructions for Use include information and
warnings which must be heeded by the user to
assure safe operation of the instrument.
Electromagnetic compatibility (EMC)
Emitted interference
Standards met:
EN 55011 (class B), EN 55022 (class B),
EN 50081-1/2
Immunity to interference
Standards met:
IEC801-2/IEC1000-4-2 (class 4), IEC801-3/
IEC1000-4-3 (class 3), IEC801-4/IEC1000-4-4
(class 4), IEC801-5/IEC1000-4-5 (class 2/3),
IEC801-6/IEC1000-4-6 (class 3), EN50082-2,
EN61000-3-2/3/IEC1000-3-2/3, EN50093/
IEC1000-4-11
Ambient temperature
Nominal operating range
+5
…
+45°C
(at 20
…
80 % atmospheric humidity)
Storage, transport
–20
…
+70°C
Diagnostic tests
Self-diagnostic test
Automatic self-diagnosis when instrument
switched on
User diagnostic test
Built-in diagnostic program (see section 5.4)
Housing
Material of cover
Polyurethane rigid foam (PUR) with fire protection
for fire class UL94VO, FCH-free
Material of base
Steel, enameled
Width
255 mm
Height
130 mm
Depth
343 mm
Weight
8.2 kg (with detector block, without accessories)