5 Notes – Maintenance – Faults

732 IC Detector / 733 IC Separation Center

144

If the rotor has been inserted correctly, its sealing surface will

be about 4 mm inside the holder. If this is not the case, bring

the rotor into the correct position from below with the aid of a

sharp object (e.g. a screwdriver).

5 Clean connection piece

Clean the sealing surface of new connection piece 98

98

(6.2832.010) with the aid of a lint-free cloth and ethanol.

6 Insert connection piece

Insert new connection piece 98

98 in suppressor holder 100

100 in

such a way that connection "1" is at the top and that the three

lugs on the connection piece fit in the corresponding open-

ings of the holder.

Screw nut 97

97 onto the thread of suppressor holder 100

100

manually (do not use tools).

7 Connect and condition the suppressor

Reconnect the suppressor to the IC system.

Before switching the suppressor to the next position for the

first time, rinse all 3 suppressor units with solution for 5 min.