2.6 Connection of 709 IC Pump

732 IC Detector / 733 IC Separation Center

21

2.6

Connection of 709 IC Pump

2.6.1

Electrical connection

For operation of the 732 IC Detector and 733 IC Separation Center you

can use any commercial HPLC pump. However, as the attainable sen-

sitivity depends to a large extent on the quality of the pump, Metrohm

advises use of the 709 IC Pump, which has been specially developed

for the demands of ion chromatography and has minimal pump pulsa-

tion and an outstanding flow constancy.
Startup and operation of the 709 IC Pump are described in the 709

Instructions for Use. The eluent, which must be degassed and filtered

(cf. section 5.1.3), is selected on the basis of the separating column in-

stalled in the 733 IC Separation Center and the current separation

problem (see 8.732.2003 IC Monograph).
The connection of the 709 IC Pump at the connection 15

15 of the 732

IC Detector is shown in Fig. 10. For this you can use the 6.2125.060

Cable available as an option or another RS cable specified as a "null

modem" cable. To ensure proper functioning of the communication

between the 732 IC Detector and 709 IC Pump, the sliding switch 36

36 on

the IC pump must be set to "RS 232" and the external control switched

on with key 88 (see 709 Instructions for Use).

6.2125.060

709

732

Fig. 10: Connection of 709 IC Pump

2.6.2

Pulsation dampener

To protect the column material against pressure drops caused by the

injector, the use of a pulsation dampener connected between the pump

and the injection valve of the 733 IC Separation Center is recom-

mended. The optional 6.2620.150 Pulsation dampener MF (see sec-

tion 7.3.1) is very well suited to this purpose.
The metal-free 6.2620.150 Pulsation Dampener is supplied fully as-

sembled and has two connections for capillaries, for which either the

connectors supplied or two 6.2744.010 PEEK compression fittings can

be used. The flow direction is arbitrary. The pulsation dampener is po-

sitioned in the interior of the 733 IC Separation Center on the base be-

low the injection valve (see Fig. 16 and Fig. 17).