4.9 Examples of methods

732 IC Detector / 733 IC Separation Center

129

4.9.2

Anion determination with Metrosep Anion Dual 2

This example shows a method for the determination of monovalent and

divalent anions with the 6.1006.100 IC Anion Column Metrosep Anion

Dual 2 and the Metrohm suppressor module, which can be used with or

without an autosampler.

General conditions

Separating column:

6.1006.100 Metrosep Anion Dual 2

Sample loop:

20

µ

L

Eluent:

1.3 mmol/L sodium carbonate,

2.0 mmol/L sodium bicarbonate

Flow:

0.8 mL/min

Eluent conductivity:

ca. 14

µ

S/cm

Full scale:

50

µµS/cm and 5 µµS/cm

Suppressor module:

Regeneration solution 20 mmol/L H

2

SO

4

,

rinsing solution Milli-Q water

Configuration

CONFIG
>CONFIG/detector
thermostat: 35 °C
'zero' unit:

µ

S/cm

cell constant 17.1 /cm
>CONFIG/printer
id.1
id.2
print header: once
date&time: on
send to: IBM
>CONFIG/print meas.value
print crit.: off
>CONFIG/auxiliaries
run number 0
number of cycles 1
>CONFIG/aux/event
enabled in program: off
date 1995-10-16
time 10:59:20
dialog: english
device label
program 732.0012
>CONFIG/aux/beep
status: on
repeat time 60 s
>CONFIG/RS settings
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: HWs
control via RS: on
>CONFIG/RS settings 709
baud rate: 9600
>CONFIG/733 IC Sep.Cent.
valve A + suppressor
control: no restriction
>CONFIG/733/valve A
status inject
trigger: inject
>CONFIG/733/suppressor
auto step: fill
status in position
trigger: ---