6.2 Remote interfaces

732 IC Detector / 733 IC Separation Center

183

6.2 Remote interfaces

The 732 IC Detector has the two remote interfaces 14

14 "733 IC Separa-

tion Center" and 17

17 "Remote" (see Fig. 3) which each have 8 input and

8 output lines.

6.2.1 "Remote" interface

The remote interface 17

17 "Remote" is used for the connection of any

type of external device (see section 2.9.8) and has the following pin as-

signment:

732 IC Detector

Pin

Function

Inputs

+5V

21

Print

Triggering the print command
Equivalent to pressing the key (see section 4.8.1).

H

L

t

P

> 100 ms

+5V

+5V

+5V

+5V

9

22

10

23

2

0

2

1

2

2

2

3

Triggering of functions (remote control)
Assignment of the input lines see section 4.4.1

Input

Decimal

Function

23

10

22

9

code

(default settings)

H

H

H

H

0

Inactive

H

H

H

L

1

PROG R/S

H

H

L

H

2

PUMP R/S

H

H

L

L

3

FILL B/STEP

H

L

H

H

4

FILL A

H

L

H

L

5

REPORT

H

L

L

H

6

ZERO

H

L

L

L

7

MARK

L

H

H

H

8

INJECT A

L

H

H

L

9

ZERO OFF

L

H

L

H

10

PLOT

L

H

L

L

11

Switch polarity

L

L

H

H

12

INJECT B

L

L

H

L

13

SELECT

L

L

L

H

14

QUIT

L

L

L

L

15

ENTER

H

L

t

P

> 100 ms

+5V

11
24
12

not assigned