beautypg.com

Metrohm 732 IC Detector User Manual

Page 94

background image

4.4 Basic settings

732 IC Detector / 733 IC Separation Center

87

>SETUP
>SETUP/peripherals

Setting the peripherals

>SETUP/peripherals
operation with 733: on

on, off

Operation with 733 IC Separation Center

on

The 732 IC Detector is operated with the 733

IC Separation Center (normal case).

Off

The 732 IC Detector is operated without the

733 IC Separation Center. All parameters

which affect the 733 IC Separation Center are

suppressed in the dialog.

>SETUP/peripherals
operation with 709: on

on, off

Operation with 709 IC Pump

on

The 732 IC Detector is operated with the 709

IC Pump (normal case).

off

The 732 IC Detector is operated without the

709 IC Pump. All parameters which affect the

709 IC Pump are suppressed in the dialog. In

this case, the RS interface "709 IC Pump" can

be used for the connection of an external

printer.