beautypg.com

5 sample determination – Metrohm 732 IC Detector User Manual

Page 69

background image

3 Operating tutorial

732 IC Detector / 733 IC Separation Center

62

3.5

Sample determination

Following the calibration of the IC system as described in section 3.4

the first sample solution can be injected.

1 Filter drinking water sample

Filter the drinking water sample with a 0.45

µ

m microfilter.

2 Trigger auto-zero

ZERO

Press the key. This starts the automatic elec-

tronic background compensation. During zeroing, the

green LED in the key starts to flash. As soon as zeroing is

at an end, the green LED in the key is perma-

nently on.

3 Set injection valve A to "FILL"

FILL

Press the key for injection valve A on the 733 IC

Separation Center. This switches the injection valve to the

"FILL" position. This position is indicated by lighting up of

the green LED in the key.

4 Fill sample loop

Immerse the suction tubing 88

88 attached to connection 22

22

in the vessel containing the drinking water sample.

Use the syringe fastened to connection 21

21 to siphon in ca.

1 mL sample.

5 Set injection valve A to "INJECT"

INJECT

Press the key for injection valve A on the 733 IC

Separation Center. This switches the injection valve to the

"INJECT" position and the chromatogram is started. The

"INJECT" position is indicated by lighting up of the green

LED in the key.

At the same time start the recorder (if used).
An integrator attached to the 733 IC Separation Center or a

PC with evaluation software (e.g. «IC Metrodata for Win95»)

is started automatically when injection valve A is switched

to the "INJECT" position.

+0.000

µµS/cm 0.0 min

full scale: 4.00

µµS/cm

As soon as injection valve A is in the "INJECT" position, the

current time is reset to

"

0.0 min

"

. This marks the start time

of the chromatogram.
The ion chromatogram of the drinking water sample is

recorded during the next 20 min or so.
Fig. 20 shows an example of a drinking water sample.