beautypg.com

Metrohm 732 IC Detector User Manual

Page 176

background image

6.1 RS232 interfaces

732 IC Detector / 733 IC Separation Center

169

Object

Meaning

Entry range/Selection

Ã.

P

arameter

General program parameters

³

Ã

.

T

ype

Program type: cycle, remote, inject

cycle

, remote, inject

³

Ã

.

C

ycle

Loop program

³

³

À

.

N

o

Number of cycles

1

999

³

À

.

S

tatus

Program status: inactive, active, test

disabled

, enabled, test

³

Ã.

E

dit

Edit program

³

À

.

S

elect

Selection of program steps (max. 20)

³

Ã

.

1

Program step 1

³

³

Ã

.

T

ime

Time for program step 1

x.x

, 0.0

999.9 min

³

³

À

.

A

ction

Actions of program step 1

³

³

Ã

.

F

lag

Program flag

---

, return, reset, end

³

³

Ã

.

FS

Change

Change full-scale range

---

, on

³

³

Ã

.

R

ange

³

³

³

Reset measurement range

100

, 200, 500

µ

S/cm,

1, 2, 5, 10 mS/cm

³

³

Ã

.

Fu

llScale

³

³

³

Reset full-scale range

0.05

100

999

µ

S/cm,

1

10 mS/cm

³

³

Ã

.

Z

ero

On/off switching of auto-zero

---

, on, off

³

³

Ã

.

P

olarity

Change polarity

---

, +, –

³

³

Ã

.

M

ark

Trigger marking signal

---

, on

³

³

Ã

.

V

alveA

Switch injection valve A

---

, fill, inject

³

³

Ã

.

ValveB

Switch injection valve B

---

, fill, inject

³

³

Ã

.

S

uppressor

Switch suppressor module

---

, step

³

³

Ã

.

Re

mote

³

³

³

³

³

³

³

³

³

Set remote output lines 1

8:

T

(leave unchanged), 0 (off, inactive, open),

1 (on, active, 0 V)

V

, 0, 1

³

³

Ã

.

Pu

mpRS

On/off switching of pump drive

---

, on, off

³

³

Ã

.

Flo

w

Change flow rate

---

, 0.01

5.00 mL/min

³

³

Ã

.

Pm

ax

Change maximum shutoff pressure

---

, 0.1

50 MPa

³

³

À

.

Pmi

n

Change minimum shutoff pressure

---

, 0.0

50 MPa

³

:

³

À

.

20

Program step 20

³

Ã

.

T

ime

Time for program step 20

x.x

, 0.0

999.9 min

³

À

.

A

ction

Actions of program step 20

³

:

see Program step 1

³

À.

D

eleteAll

Delete all program steps

possible only in program status "disabled"

$G

&S

electEvent

Events

³

Ã.

1

Event 1

³

Ã

.

D

ate

Date/time or only time (daily) for event 1

YY-MM-DD HH:MM:SS

³

À

.

A

ction

Actions of event 1 1

³

Ã

.

F

SChange

Change full-scale range

---

, on

³

Ã

.

R

ange

³

³

Reset measurement range

100

, 200, 500

µ

S/cm,

1, 2, 5, 10 mS/cm

³

Ã

.

Fu

llScale

³

³

Reset full-scale range

0.05

100

999

µ

S/cm,

1

10 mS/cm

³

Ã

.

Z

ero

On/off switching of auto-zero

---

, on, off

³

Ã

.

P

olarity

Change polarity

---

, +, –