Metrohm 732 IC Detector User Manual

Page 27

background image

2 Installation

732 IC Detector / 733 IC Separation Center

20

49

49

Ferrule (6.2620.010)

50

50

Pressure screw (6.2620.000)

51

51

Capillary

6.2620.020 Steel capillary or

6.1831.010 PEEK capillary

52

52

Compression fitting

(6.2744.010)

2.5.3

PEEK connectors

For the connection of 6.1822.010 PEEK capillaries (i.D. = 0.3 mm) or

6.1822.010 PTFE microcapillaries, the 6.2744.010 PEEK compres-

sion fittings are used. Proceed as follows:

1 Mount compression fitting

Slide a compression fitting 52

52 (6.2744.010) over the end of the

capillary 51

51 to be fastened as shown in Fig. 9.

2 Insert capillary in connection

Push capillary end in the corresponding connection as far as it

will go (to avoid dead volume).

3 Tighten compression fitting

Tighten compression fitting 52

52 by hand (never use tools).

50

50

49

49

51

51

52

52

Fig. 9:

Connectors for capillaries