4.9 Examples of methods

732 IC Detector / 733 IC Separation Center

131

Settings for 709 IC Pump

Installation:

as shown in Fig. 10 (see section 2.6)

key:

Switched on (remote operation activated)

Settings for 752 Pump Unit

Installation:

as described in section 2.8.6

Chromatogram for anion standard

FULL REPORT

Ret Time Component Concentr. Area Height

(Min) Name ppm (uV*Sec) (uV)

-------- --------------- ----------- ---------- ----------

4.918 Chloride 19.661419 4187692.25 331573.781

11.505 Nitrate 2.010445 1906272.75 68276.7421

13.399 Phosphate 1.973833 1011560.75 30180.7343

16.617 Sulphate 19.895790 2717087.00 69003.4375

Fig. 23: Ion chromatogram for anion stan-

dard with Metrosep Anion Dual 2

Full Scale

5

µ

S/cm

Cl

SO

4

2–

HPO

4

2–

NO

3