beautypg.com

Metrohm 732 IC Detector User Manual

Page 37

background image

2 Installation

732 IC Detector / 733 IC Separation Center

30

2.7.3

Precolumns with cartridge head

The precolumn cartridges are installed in the 6.2821.040 Cartridge

head as follows (see Fig. 15):

1 Prepare separating column

Remove end caps from separating column 76

76.

Unscrew fastening screw from column inlet.

Take steel connector 79

79 for ferrule out of fastening screw.

2 Install cartridge

Remove end caps for precolumn cartridge 74

74 (the steel mesh

73

73 and gaskets 59

59 are already installed in the cartridge).

Mount steel spacer 78

78 on separating column 76

76.

Mount precolumn cartridge 74

74 on the steel spacer (comply

with flow direction if specified on the precolumn).

Screw Manufit pressure screw 77

77 firmly to separating column

76

76.

Mount steel connector 79

79 for ferrule on the inlet side of the

precolumn cartridge 74

74.

Screw on Manufit housing 80

80 with Manufit pressure screw 77

77.

3 Connect precolumn

Fit connectors to the column connection capillary 67

67

mounted on the injection valve (see section 2.5).

Screw column connection capillary 67

67 to Manufit housing 80

80.