background image

4.4 Basic settings

732 IC Detector / 733 IC Separation Center

89

When the heating is switched on, the green display

lamp 77 "THERMOSTAT" is on. Thermostatting of the

conductivity detector is switched off by pressing the

key. The display shows the value

"

off °C

"

,

at the same time the green display lamp 77 goes out.

Thermostatting functions only if the ambient tempera-

ture is at least 5 °C lower than the operating tempera-

ture.

>CONFIG/detector
‘zero’ unit:

µµS/cm

µ

S/cm, %fs, mV

Unit for display of the auto-zero value
The following units can be selected for the auto-zero

value displayed in the instrument standby mode (see

section 4.6.3):

µ

S/cm

Display in conductivity unit

µ

S/cm.

%fs

Display in % of the full-scale range which

has been set under the key

(see section 4.5.1) or

key (see section 4.5.2).

mV

Display in mV.

>CONFIG/detector
cell constant 16.7/cm

13.0

21.0 /cm

Cell constant of the conductivity cell
Setting of the cell constant of the measuring cell for

correct display of the absolute conductivity.
In ion chromatography, interest is normally centered

only on relative changes in the conductivity and not on

the absolute value. With the cell constant of 16.7/cm set

in the factory, the error in the display of the absolute

conductivity is ca.

±

10%.

If you wish to have a more accurate display of the ab-

solute conductivity, the cell constant must be deter-

mined by means of a calibration solution. To do this,

pump a solution of known conductivity through the IC

system, observe the display of the absolute conductivity

and change the cell constant until the displayed value

matches the actual value.

>CONFIG
>CONFIG/printer

Settings for external printer

>CONFIG/printer
id.1

18 ASCII characters

Identification 1 for 1

st

line of print header

Freely selectable ASCII character string for the first line

of the print header (text entry, section 4.2.5).