2.8 Separating columns and suppressor module
732 IC Detector / 733 IC Separation Center
35
2.8.4
One-channel system without suppressor module
With the one-channel system without suppressor module, the IC sepa-
rating column is installed in the 2.733.0010 IC Separation Center as
follows (see Fig. 16):
1 Connect column to injector
•
Remove end caps from column 76
76.
•
without precolumn:
Screw inlet end of separating column 76
76 (note flow direction)
to column connection capillary 67
67 mounted on the injector.
•
with precolumn in twin cartridge holder:
Screw inlet end of separating column 76
76 (note flow direction)
to outlet capillary 72
72 (see Fig. 15).
•
with precolumn in cartridge head:
Install separating column 76
76 (note flow direction) in the car-
tridge head (see Fig. 15) as described in section 2.7.3.
2 Rinse column
•
Place a beaker beneath the column outlet.
•
Switch on 709 IC Pump and rinse the column for ca. 10 min
with eluent.
•
Switch off IC Pump.
3 Connect column to detector block
•
Screw outlet end of separating column 76
76 to the inlet capillary
82
82 permanently mounted on the detector block.
4 Fix column
•
Insert one or two column holders 84
84 (6.2027.030, 6.2027.040
or 6.2027.050) in the mounting rails 83
83 and fasten separating
column in the column holder.
2.8.5
Two-channel system without suppressor module
With the two-channel system without suppressor module (2.733.0X20
IC Separation Center), the first IC separating column is connected on
the left side to injection valve A and detector block A as with the one-
channel system (see section 2.8.4 and Fig. 16). The second column is
connected on the right side to injection valve B and detector block B in
an analogous manner.