2 Installation

732 IC Detector / 733 IC Separation Center

40

2.8.7

Leak testing and conditioning

Before sample solutions can be injected in the IC system, the entire

system must be tested for leaks and then conditioned with eluent until

the baseline is stable. Proceed as follows:

1 Switch on 709 IC Pump

Immerse aspirating capillary of the 709 IC Pump in eluent.

Set flow rate recommended on the 709 IC Pump for the

separating column used (normally 0.5

2 mL/min).

Set maximum shutoff pressure on the 709 IC Pump

(normally ca. 3 MPa above the pressure observed with the

column used).

Switch on 709 IC Pump.

2 Check for leaks

Check all capillaries and their connections between the 709

IC Pump and the detector block for escaping liquid. If eluent

escapes anywhere, the appropriate compression fitting must

be tightened further or changed.

3 Switch on 732 IC Detector

Switch on 732 IC Detector with mains switch 99.

Set operating temperature: Under the key,

enter the desired value for the parameter "

thermostat

"

(default value = 35 °C).

Enter cell constant: Under the key, enter the

value printed on the detector block for the parameter "

cell

constant

".

Set measuring range: Under the key, set the

parameter "

range

" so that the displayed absolute conductivity

value of the eluent lies within the selected range (default

value = 1 mS/cm).

Set full-scale range: Under the or

SCALE> key, enter the desired value for the parameter "

full

scale

" (default value = range/1). To start with, it is advisable

not to set too narrow a full-scale range as the conductivity of

the eluent in the conditioning phase can change greatly until

attainment of a stable temperature.

4 Condition IC system

Rinse the IC system with eluent until the desired stability of

the baseline is reached (normally 30

60 min; if the eluent is

changed, the establishment of the ion exchanger equilibrium

on the separating column can take longer).

5 Condition suppressor (if present)

Load method "

Prep-MSM

" and start this method. The suppres-

sor module is switched to the next position every 20 min and

in this way conditioned.