2 Installation

732 IC Detector / 733 IC Separation Center

14

3 Connect waste container

Lead the outlet capillary of the detector block to a sufficiently

large waste container and fix in place.

4 Connect 732 to 733

Plug one end of the 6.2125.090 Connecting cable into con-

nection 14

14 "733 IC Separation Center” of the 732 IC Detector

and fasten to the instrument by tightening the screws in the

cable connector (see Fig. 6).

Plug the other end of the 6.2125.090 Connecting cable into

connection 36

36 "732 IC Detector" of the 733 IC Separation

Center and fasten to the instrument by tightening the screws

in the cable connector (see Fig. 6).

2.3.2

2.733.0X20 IC Separation Center

For operation of the 2.733.0020 or the metal-free 2.733.0120 instrument

version of the 733 Separation Center, two 1.732.0100 or metal-free

1.732.0110 IC Detectors are needed. It is best to proceed as follows

when connecting the instruments and the two detector blocks:

1 Install detector blocks

Unscrew the four knurled screws 34

34 from the top rear panel

35

35 of the 733 IC Separation Center and remove rear panel

(see Fig. 5).

Position first detector block A from the back in the space

provided in the 733 IC Separation Center on the right and

push fully to the front (see Fig. 16).

Position second detector block B from the back in the space

provided in the 733 IC Separation Center on the left and push

fully to the front (see Fig. 16).

Insert the cable permanently attached to detector block A in

opening 33

33 and the outlet capillary in opening 32

32 "Waste A"

of the rear panel 35

35.

Insert the cable permanently attached to detector block B in

opening 29

29 and the outlet capillary in opening 30

30 "Waste B"

of the rear panel 35

35.

Replace rear panel 35

35 and fasten to 733 IC Separation Center

with the four knurled screws 34

34.

2 Connect detector blocks

Plug the gray connecting cable permanently attached to

detector block A into connection 13

13 "Detector block" of the

first 732 IC Detector and fasten to the instrument by tighten-

ing the screws in the cable connector (see Fig. 7).

Plug the gray connecting cable permanently attached to

detector block B into connection 13

13 "Detector Block" of the

second 732 IC Detector Block and fasten to the instrument by

tightening the screws in the cable connector.