4 Operation
732 IC Detector / 733 IC Separation Center
110
program step can be edited.
Change time for existing program steps
For a program step selected with
time can be entered directly with the numeric keys and
must be confirmed with
program points of the program step can then be edited
(if no change is necessary, press
Delete existing program steps
A program step selected with
deleted by pressing the two keys
program steps at once, it is better to use the submenu
"
PROGRAM/delete all
"
.
>PROGRAM/edit. XXX.X min
XXXXXXXXX ---
Edit program points
After confirmation of the newly entered or selected time
with
possible program points of a program step. The time of
the program step then appears in the top line of the
display, whereas in the bottom line the program points
listed below associated with the program step can be
edited in turn. For all program points, the default value
"
---
"
means that no action results.
>PROGRAM/edit. XXX.X min
flag: ---
---, return, reset, end
Set program flag
After setting of a program flag, no additional program
points can be entered.
return
Return with loop programs
Each time the
"
return
"
flag is reached, a
return is made to the start of the program un-
til the inputted number of cycles has been
reached.
reset
Reset the parameters
With this flag all parameters of the
key are reset to the values valid
before the start of the program.
end
Program end
This flag is used to end a program, additional
program steps are no longer executed
.
Pro-
grams without
"
end
"
are not stopped until
>PROGRAM/edit. XXX.X min
FS change: ---
---, on
Change full-scale range
on
The full-scale range should be changed. As
the possible settings of the full-scale range
depend on the measuring range, both the
measuring range and the full-scale range