4 Operation

732 IC Detector / 733 IC Separation Center

110

program step can be edited.
Change time for existing program steps

For a program step selected with can be

deleted by pressing the two keys and

simultaneously. If you wish to delete all

program steps at once, it is better to use the submenu

"

PROGRAM/delete all

"

.

>PROGRAM/edit. XXX.X min
XXXXXXXXX ---

Edit program points
After confirmation of the newly entered or selected time

with , it is possible to edit the maximum 14

possible program points of a program step. The time of

the program step then appears in the top line of the

display, whereas in the bottom line the program points

listed below associated with the program step can be

edited in turn. For all program points, the default value

"

---

"

means that no action results.

>PROGRAM/edit. XXX.X min
flag: ---

---, return, reset, end

Set program flag
After setting of a program flag, no additional program

points can be entered.

return

Return with loop programs

Each time the

"

return

"

flag is reached, a

return is made to the start of the program un-

til the inputted number of cycles has been

reached.

reset

Reset the parameters

With this flag all parameters of the

key are reset to the values valid

before the start of the program.

end

Program end

This flag is used to end a program, additional

program steps are no longer executed

.

Pro-

grams without

"

end

"

are not stopped until

is pressed.

>PROGRAM/edit. XXX.X min
FS change: ---

---, on

Change full-scale range

on

The full-scale range should be changed. As

the possible settings of the full-scale range

depend on the measuring range, both the

measuring range and the full-scale range

must be reentered.