beautypg.com

Metrohm 732 IC Detector User Manual

Page 116

background image

4.7 Programming

732 IC Detector / 733 IC Separation Center

109

test

Program test

The test status is used to test a program. Af-

ter the program status has been set to

"

test

"

and return to the standby mode, a

switch is automatically made to the test

mode, which runs as follows:

The bottom line of display shows the

message

"

test step

"

with details of the

time

"

X.XXX min

"

when the first program

step should be executed.

Pressing the key executes the

actions associated with the program

step.

The display shows the next program

steps in turn which are triggered with

.

After execution of the last program step,

the message

"

program end

"

appears

and must be confirmed with .

The instrument then returns to the

standby mode and the program status is

set to

"

disabled

"

.

>PROGRAM
>PROGRAM/edit

Edit program
This submenu is used for the entry of new program steps or

to edit existing program steps and their program points.

>PROGRAM/edit
time: x.x min

0.0

999.9 min

Edit program step
Under this inquiry new program steps can be entered or

existing ones selected, changed and deleted. Each

program step includes a definite time (two program

steps with the same time can not be entered). A pro-

gram can contain up to 20 program steps.

Enter new program step

A new program step can be entered when

"

x.x min

"

for the time appears in the display. The new time en-

tered with the numeric keys must be confirmed with the

key, the individual program points which

should be executed at this time can then be entered.

Select existing program steps

Program steps already entered can be selected with the