Metrohm 732 IC Detector User Manual
Page 116
4.7 Programming
732 IC Detector / 733 IC Separation Center
109
test
Program test
The test status is used to test a program. Af-
ter the program status has been set to
"
test
"
and return to the standby mode, a
switch is automatically made to the test
mode, which runs as follows:
•
The bottom line of display shows the
message
"
test step
"
with details of the
time
"
X.XXX min
"
when the first program
step should be executed.
•
Pressing the
actions associated with the program
step.
•
The display shows the next program
steps in turn which are triggered with
•
After execution of the last program step,
the message
"
program end
"
appears
and must be confirmed with
The instrument then returns to the
standby mode and the program status is
set to
"
disabled
"
.
>PROGRAM
>PROGRAM/edit
Edit program
This submenu is used for the entry of new program steps or
to edit existing program steps and their program points.
>PROGRAM/edit
time: x.x min
0.0
…
999.9 min
Edit program step
Under this inquiry new program steps can be entered or
existing ones selected, changed and deleted. Each
program step includes a definite time (two program
steps with the same time can not be entered). A pro-
gram can contain up to 20 program steps.
Enter new program step
A new program step can be entered when
"
x.x min
"
for the time appears in the display. The new time en-
tered with the numeric keys must be confirmed with the
should be executed at this time can then be entered.
Select existing program steps
Program steps already entered can be selected with the
pressed, the individual program points of the selected