3 faults and malfunctions – Metrohm 732 IC Detector User Manual
Page 154
5.3 Faults and malfunctions
732 IC Detector / 733 IC Separation Center
147
Error message
Meaning/Causes
Rectification
range error
#
Error in internal hardware test.
Inform Metrohm service.
RS error 36
RS receive error; wrong parity.
devices.
RS error 37
RS receive error; wrong stop bit.
devices.
RS error 38
RS-receive error; overflow (at least 1
character could not be read).
devices, restart sender.
RS error 39
RS receive error; overflow of internal
receive buffer (>82 characters).
RS error 40
RS transmission error; DSR = OFF.
Handshake unsatisfactory for more
than 1 s.
and ready?).
RS error 41
RS transmission error; DCD = ON.
Handshake unsatisfactory for more
than 1 s.
and ready?).
RS error 42
RS transmission error; CTS = OFF.
Handshake unsatisfactory for more
than 1 s.
and ready?).
RS error 43
RS transmission error; transmission
of the IC Detector interrupted with
XOFF for at least 3 s.
RS error 44
RS transmission error; the RS232
parameters for the two devices are
no longer the same.
both devices.
RS error 45
RS transmission error; the receive
buffer of the IC Detector contains an
incomplete string (L
F
missing), the
transmission is thus blocked.
F
.
RS error 60
…
89
Error in RS232 test.
two RS232 interfaces.
RS error XX at 709
RS error 36
…
89 at the RS232
interface "709 IC Pump".
See individual RS errors.
RS error XX at RS 232
RS error 36
…
89 at the RS232
interface "RS 232".
See individual RS errors.
sensitivity error
#
Error in internal hardware test.
Inform Metrohm service.
stopped:YY-MM-DD HH:MM
709 IC Pump stopped at specified
time owing to violation of shutoff
limits.
Press the 709 IC Pump keys
again.
store operation failed
Storage of the method failed.
Switch instrument off/on. Store method
again. If error reappears, inform
Metrohm service.
U cell error
#
Error in internal hardware test.
Inform Metrohm service.