beautypg.com

3 faults and malfunctions – Metrohm 732 IC Detector User Manual

Page 154

background image

5.3 Faults and malfunctions

732 IC Detector / 733 IC Separation Center

147

Error message

Meaning/Causes

Rectification

range error

#

Error in internal hardware test.

Inform Metrohm service.

RS error 36

RS receive error; wrong parity.

, set same parity for both

devices.

RS error 37

RS receive error; wrong stop bit.

, set same stop bit for both

devices.

RS error 38

RS-receive error; overflow (at least 1

character could not be read).

, set same baud rate for both

devices, restart sender.

RS error 39

RS receive error; overflow of internal

receive buffer (>82 characters).

.

RS error 40

RS transmission error; DSR = OFF.

Handshake unsatisfactory for more

than 1 s.

, check receiver (switched on

and ready?).

RS error 41

RS transmission error; DCD = ON.

Handshake unsatisfactory for more

than 1 s.

, check receiver (switched on

and ready?).

RS error 42

RS transmission error; CTS = OFF.

Handshake unsatisfactory for more

than 1 s.

, check receiver (switched on

and ready?).

RS error 43

RS transmission error; transmission

of the IC Detector interrupted with

XOFF for at least 3 s.

or send XON.

RS error 44

RS transmission error; the RS232

parameters for the two devices are

no longer the same.

, reset RS232 parameters for

both devices.

RS error 45

RS transmission error; the receive

buffer of the IC Detector contains an

incomplete string (L

F

missing), the

transmission is thus blocked.

or send L

F

.

RS error 60

89

Error in RS232 test.

, check connection between the

two RS232 interfaces.

RS error XX at 709

RS error 36

89 at the RS232

interface "709 IC Pump".

See individual RS errors.

RS error XX at RS 232

RS error 36

89 at the RS232

interface "RS 232".

See individual RS errors.

sensitivity error

#

Error in internal hardware test.

Inform Metrohm service.

stopped:YY-MM-DD HH:MM

709 IC Pump stopped at specified

time owing to violation of shutoff

limits.

Press the 709 IC Pump keys ,

in succession and then

again.

store operation failed

Storage of the method failed.

Switch instrument off/on. Store method

again. If error reappears, inform

Metrohm service.

U cell error

#

Error in internal hardware test.

Inform Metrohm service.