beautypg.com

4 diagnostic tests, General, Preparing instruments – Metrohm 732 IC Detector User Manual

Page 156: Checking working memory (ram), Check keypad, Check display, Check rs232 interface, Check remote interface, Internal hardware test, Initialize data memory

background image

5.4 Diagnostic tests

732 IC Detector / 733 IC Separation Center

149

5.4

Diagnostic tests

5.4.1

General

The 732 IC Detector and 733 IC Separation Center are very precise and

dependable instruments. Thanks to their rugged construction, it is

highly unlikely that external mechanical or electrical influences will have

any adverse effect on their functions.
Although a fault in the instruments can not be excluded with certainty,

the possibility is greater that malfunctions are caused by wrong opera-

tion or handling, through improper connections and the operation with

third-party devices.
Whatever the case, it is always advisable to localize the fault with the

diagnostic tests, which can be performed quickly and simply. The cus-

tomer need call Metrohm service only when the instrument really has a

fault. Further, he can use the results of the specific diagnostic function

to provide the service engineer with much more precise information.
In the case of inquiries, always quote the serial number on model plate

10

10 of the 732 IC Detector (see Fig. 3) or on model plate 47

47 of the 733

IC Separation Center (see Fig. 5), program version (see section 4.4.2)

and if applicable the error message.

Procedure
The diagnostic menu listed in section 5.4.2 shows all components for

which detailed instructions (diagnostic steps) are available for checking

the functionality.
In the case of a possible malfunction, we advise you to perform either

the corresponding diagnostic step or all diagnostic steps as a routine

step on the instrument.
The reactions of the instruments to the instructions must be compared

with the descriptions in the diagnostic step. If the instruments do not

show the expected reaction ("No" case), the appropriate diagnostic

step must be repeated to exclude operating errors. However, it is highly

probable that repeated wrong reactions indicate a malfunction.

Instruments needed:
Necessary only if RS232 or remote should be checked:

3.496.8510 Test plug (at ‘"Remote" socket)

3.496.8480 Test plug (at "RS 232" socket)