6 Interfaces

732 IC Detector / 733 IC Separation Center

188

6.5

Valve interfaces

The two valve interfaces 38

38 and 46

46 of the 733 IC Separation Center can

be used for the attachment of external devices (e.g. 750 Autosampler,
«

IC Metrodata for Win95») which can be employed to control the injec-

tion valves or the suppressor module (see section 2.3). The two valve

interfaces for the injection valves A and B (or the suppressor module)

are identical and have the following connections

733 IC Sep. Center Connection

Function

Inputs

0V

Ground

Ground

+5V

10k

Fill

Valve

→ "FILL"

Valve is switched to the position "FILL"
(or suppressor module is switched further)

H

L

t

P

> 2 ms

+5V

10k

Inject

Valve

→ "INJECT"

Valve is switched to the position "INJECT"

H

L

t

P

> 2 ms

Outputs

COM

RUN

Position "Fill"
A pulse is outputted when the valve is switched to
the position "FILL".

H

L

t

P

≅≅

250 ms

COM

RUN

Integrator Start
A pulse is outputted when the valve is switched to
the position "INJECT".

H

L

t

P

≅≅

250 ms

No liability whatsoever will be accepted for damage arising from the improper connection of devices.