5 Notes – Maintenance – Faults
732 IC Detector / 733 IC Separation Center
152
Key table
Code
Key
Code
Key
0
<9 / METHOD>
12
1
<8 / PUMP R/S>
13
2
<7 / CONFIG>
14
<3 / PRINT>
3
15
<2 / REPORT>
4
16
<1 / PLOT>
5
17
6
<6 / PROGRAM>
18
7
<5 / EVENT>
19
8
<4 / ZERO OFF>
20
<-/+ / è>
9
21
<. / ç>
10
22
<0 / MARK>
11
23
5.4.5
Check display
This diagnostic step is used to check all light emitting diodes and the
display for functionality.
•
Prepare instrument for diagnostic test (see section 5.4.2).
•
If need be, press <9> key repeatedly until
diagnosis
>diag/display test
•
After the
test sequence for a visual check on the light emitting diodes and the display.
⇒
The light emitting diodes for OVERLOAD, THERMOSTAT, ZERO and
PROG R/S are switched on for a certain time.
⇒ Power on test pattern (each pixel active) appears.
⇒
Both lines of the display are cleared.
⇒
Both lines of the display are written to with the characters “#”, “H”
and finally with “I”.
⇒
Both lines are written to from right to left with the continuous moving
display „0123456789ABCDEFGHIJKLMNOPQRSTUVWXYZ“.
⇒ The light emitting diodes for OVERLOAD, THERMOSTAT, ZERO and
PROG R/S flash in succession for a brief time.
•
The test sequence can be held by pressing any key (with the excep-
tion of <9>) and restarted.
•
The test is quit with the <9> key.
diagnosis
>diag/RS test